Fluke 199B Service Manual

Download Service manual of Fluke 192B Accessory, Measuring Instruments for Free or View it Online on All-Guides.com. This version of Fluke 192B Manual compatible with such list of devices, as: 192C, 196C, 199B, 199C, 225C

Fluke 192B Service manual - Page 1
1
Fluke 192B Service manual - Page 2
2
Fluke 192B Service manual - Page 3
3
Fluke 192B Service manual - Page 4
4
Fluke 192B Service manual - Page 5
5
Fluke 192B Service manual - Page 6
6
Fluke 192B Service manual - Page 7
7
Fluke 192B Service manual - Page 8
8
Fluke 192B Service manual - Page 9
9
Fluke 192B Service manual - Page 10
10
Fluke 192B Service manual - Page 11
11
Fluke 192B Service manual - Page 12
12
Fluke 192B Service manual - Page 13
13
Fluke 192B Service manual - Page 14
14
Fluke 192B Service manual - Page 15
15
Fluke 192B Service manual - Page 16
16
Fluke 192B Service manual - Page 17
17
Fluke 192B Service manual - Page 18
18
Fluke 192B Service manual - Page 19
19
Fluke 192B Service manual - Page 20
20
Fluke 192B Service manual - Page 21
21
Fluke 192B Service manual - Page 22
22
Fluke 192B Service manual - Page 23
23
Fluke 192B Service manual - Page 24
24
Fluke 192B Service manual - Page 25
25
Fluke 192B Service manual - Page 26
26
Fluke 192B Service manual - Page 27
27
Fluke 192B Service manual - Page 28
28
Fluke 192B Service manual - Page 29
29
Fluke 192B Service manual - Page 30
30
Fluke 192B Service manual - Page 31
31
Fluke 192B Service manual - Page 32
32
Fluke 192B Service manual - Page 33
33
Fluke 192B Service manual - Page 34
34
Fluke 192B Service manual - Page 35
35
Fluke 192B Service manual - Page 36
36
Fluke 192B Service manual - Page 37
37
Fluke 192B Service manual - Page 38
38
Fluke 192B Service manual - Page 39
39
Fluke 192B Service manual - Page 40
40
Fluke 192B Service manual - Page 41
41
Fluke 192B Service manual - Page 42
42
Fluke 192B Service manual - Page 43
43
Fluke 192B Service manual - Page 44
44
Fluke 192B Service manual - Page 45
45
Fluke 192B Service manual - Page 46
46
Fluke 192B Service manual - Page 47
47
Fluke 192B Service manual - Page 48
48
Fluke 192B Service manual - Page 49
49
Fluke 192B Service manual - Page 50
50
Fluke 192B Service manual - Page 51
51
Fluke 192B Service manual - Page 52
52
Fluke 192B Service manual - Page 53
53
Fluke 192B Service manual - Page 54
54
Fluke 192B Service manual - Page 55
55
Fluke 192B Service manual - Page 56
56
Fluke 192B Service manual - Page 57
57
Fluke 192B Service manual - Page 58
58
Fluke 192B Service manual - Page 59
59
Fluke 192B Service manual - Page 60
60
Fluke 192B Service manual - Page 61
61
Fluke 192B Service manual - Page 62
62
Fluke 192B Service manual - Page 63
63
Fluke 192B Service manual - Page 64
64
Fluke 192B Service manual - Page 65
65
Fluke 192B Service manual - Page 66
66
Fluke 192B Service manual - Page 67
67
Fluke 192B Service manual - Page 68
68
Fluke 192B Service manual - Page 69
69
Fluke 192B Service manual - Page 70
70
Fluke 192B Service manual - Page 71
71
Fluke 192B Service manual - Page 72
72
Fluke 192B Service manual - Page 73
73
Fluke 192B Service manual - Page 74
74
Fluke 192B Service manual - Page 75
75
Fluke 192B Service manual - Page 76
76
Fluke 192B Service manual - Page 77
77
Fluke 192B Service manual - Page 78
78
Fluke 192B Service manual - Page 79
79
Fluke 192B Service manual - Page 80
80
Fluke 192B Service manual - Page 81
81
Fluke 192B Service manual - Page 82
82
Fluke 192B Service manual - Page 83
83
Fluke 192B Service manual - Page 84
84
Fluke 192B Service manual - Page 85
85
Fluke 192B Service manual - Page 86
86
Fluke 19xB-19xC-2x5C
Service Manual
5-6
5.4 Contrast Calibration Adjustment
After entering the calibration mode the display shows:
WarmingUp (CL 0200):IDLE (valid)
Do not press
now! If you did, turn the test tool off and on, and enter the
calibration mode again.
Proceed as follows to adjust the maximum display darkness (CL 0100), the default
contrast (CL 0110) , and the maximum display brightness (CL 0120).
1. Press three times to select maximum darkness calibration Contrast (CL 0100):
2. Press CALIBRATE . The display will show a dark test pattern, see Figure 5-2
3. Using
adjust the display to the maximum darkness at which the test pattern
is only just visible.
4. Press to return to the softkey menu.
5.
Press to select default contrast calibration Contrast (CL 0110):
6. Press CALIBRATE. The display shows the test pattern at default contrast.
7. Using
set the display to optimal (becomes default) contrast.
8. Press to return to the softkey menu.
9.
Press to select maximum brightness calibration Contrast (CL 0120):
10. Press CALIBRATE. The display shows a bright test pattern.
11. Using
adjust the display to the maximum brightness, at which the test pattern
is only just visible.
12. Press
to return to the softkey menu.
13. Now you can either
Exit, if only the Contrast had to be adjusted. Continue at Section 5.7.
or
Do the complete calibration. Press
to select the next step (WarmingUp),
and continue at Section 5.5.
Figure 5-2. Display Test Pattern