Fluke 199B Service Manual

Download Service manual of Fluke 192B Accessory, Measuring Instruments for Free or View it Online on All-Guides.com. This version of Fluke 192B Manual compatible with such list of devices, as: 192B, 192C, 196C, 199C, 215C

Fluke 192B Service manual - Page 1
1
Fluke 192B Service manual - Page 2
2
Fluke 192B Service manual - Page 3
3
Fluke 192B Service manual - Page 4
4
Fluke 192B Service manual - Page 5
5
Fluke 192B Service manual - Page 6
6
Fluke 192B Service manual - Page 7
7
Fluke 192B Service manual - Page 8
8
Fluke 192B Service manual - Page 9
9
Fluke 192B Service manual - Page 10
10
Fluke 192B Service manual - Page 11
11
Fluke 192B Service manual - Page 12
12
Fluke 192B Service manual - Page 13
13
Fluke 192B Service manual - Page 14
14
Fluke 192B Service manual - Page 15
15
Fluke 192B Service manual - Page 16
16
Fluke 192B Service manual - Page 17
17
Fluke 192B Service manual - Page 18
18
Fluke 192B Service manual - Page 19
19
Fluke 192B Service manual - Page 20
20
Fluke 192B Service manual - Page 21
21
Fluke 192B Service manual - Page 22
22
Fluke 192B Service manual - Page 23
23
Fluke 192B Service manual - Page 24
24
Fluke 192B Service manual - Page 25
25
Fluke 192B Service manual - Page 26
26
Fluke 192B Service manual - Page 27
27
Fluke 192B Service manual - Page 28
28
Fluke 192B Service manual - Page 29
29
Fluke 192B Service manual - Page 30
30
Fluke 192B Service manual - Page 31
31
Fluke 192B Service manual - Page 32
32
Fluke 192B Service manual - Page 33
33
Fluke 192B Service manual - Page 34
34
Fluke 192B Service manual - Page 35
35
Fluke 192B Service manual - Page 36
36
Fluke 192B Service manual - Page 37
37
Fluke 192B Service manual - Page 38
38
Fluke 192B Service manual - Page 39
39
Fluke 192B Service manual - Page 40
40
Fluke 192B Service manual - Page 41
41
Fluke 192B Service manual - Page 42
42
Fluke 192B Service manual - Page 43
43
Fluke 192B Service manual - Page 44
44
Fluke 192B Service manual - Page 45
45
Fluke 192B Service manual - Page 46
46
Fluke 192B Service manual - Page 47
47
Fluke 192B Service manual - Page 48
48
Fluke 192B Service manual - Page 49
49
Fluke 192B Service manual - Page 50
50
Fluke 192B Service manual - Page 51
51
Fluke 192B Service manual - Page 52
52
Fluke 192B Service manual - Page 53
53
Fluke 192B Service manual - Page 54
54
Fluke 192B Service manual - Page 55
55
Fluke 192B Service manual - Page 56
56
Fluke 192B Service manual - Page 57
57
Fluke 192B Service manual - Page 58
58
Fluke 192B Service manual - Page 59
59
Fluke 192B Service manual - Page 60
60
Fluke 192B Service manual - Page 61
61
Fluke 192B Service manual - Page 62
62
Fluke 192B Service manual - Page 63
63
Fluke 192B Service manual - Page 64
64
Fluke 192B Service manual - Page 65
65
Fluke 192B Service manual - Page 66
66
Fluke 192B Service manual - Page 67
67
Fluke 192B Service manual - Page 68
68
Fluke 192B Service manual - Page 69
69
Fluke 192B Service manual - Page 70
70
Fluke 192B Service manual - Page 71
71
Fluke 192B Service manual - Page 72
72
Fluke 192B Service manual - Page 73
73
Fluke 192B Service manual - Page 74
74
Fluke 192B Service manual - Page 75
75
Fluke 192B Service manual - Page 76
76
Fluke 192B Service manual - Page 77
77
Fluke 192B Service manual - Page 78
78
Fluke 192B Service manual - Page 79
79
Fluke 192B Service manual - Page 80
80
Fluke 192B Service manual - Page 81
81
Fluke 192B Service manual - Page 82
82
Fluke 192B Service manual - Page 83
83
Fluke 192B Service manual - Page 84
84
Fluke 192B Service manual - Page 85
85
Fluke 192B Service manual - Page 86
86
Performance Verification
4.1 Introduction 4
4-3
4.1 Introduction
Warning
Procedures in this chapter should be performed by qualified
service personnel only. To avoid electrical shock, do not
perform any servicing unless you are qualified to do so.
The ScopeMeter® test tool (referred to as test tool) should be calibrated and in operating
condition when you receive it.
The following performance tests are provided to ensure that the test tool is in a proper
operating condition. If the test tool fails any of the performance tests, calibration
adjustment (see Chapter 5) and/or repair is necessary.
The Performance Verification Procedure is based on the specifications, see Chapter 2 of
this Service Manual. The values given here are valid for ambient temperatures between
18 °C and 28 °C.
The Performance Verification Procedure is a quick way to check most of the test tool’s
specifications. Because of the highly integrated design of the test tool, it is not always
necessary to check all features separately.
4.2 Equipment Required For Verification
The primary source instrument used in the verification procedures is the Fluke 5500A. If
a 5500A is not available, you can substitute another calibrator as long as it meets the
minimum test requirements.
Fluke 5500A Multi Product Calibrator, including SC300 or SC600 Oscilloscope
Calibration Option.
Stackable Test Leads (4x), supplied with the 5500A.
50Ω Coax Cables (2x), Fluke PM9091 (1.5m) or PM9092 (0.5m).
Male BNC to Dual Female BNC adapter (1x), Fluke PM9093/001
50Ω feed through termination, Fluke PM9585.
Dual Banana Plug to Female BNC Adapter (1x), Fluke PM9081/001.
Dual Banana Jack to Male BNC Adapter (1x), Fluke PM9082/001.
TV Signal Generator, Philips PM5418, NOT required if SC600 Oscilloscope
Calibration Option is used.
75Ω Coax cable (1x), Fluke PM9075.
75Ω Feed through termination (1x), ITT-Pomona model 4119-75.
4.3 General Instructions
Follow these general instructions for all tests:
For all tests, power the test tool with the BC190 power adapter/battery charger. The
battery pack must be installed.
Allow the 5500A to satisfy its specified warm-up period.
For each test point , wait for the 5500A to settle.
Allow the test tool a minimum of 30 minutes to warm up.
One division on the LCD consists of 25 pixels ( 1 pixel = 0.04 division).